Please use this identifier to cite or link to this item:
https://ri.ufs.br/jspui/handle/riufs/22482
Document Type: | Monografia |
Title: | Estudo do perfil de feixe de fótons em LINAC |
Authors: | Silva, Rayanna Vieira da |
Issue Date: | 2-May-2025 |
Advisor: | Lalic, Susana de Souza |
Resumo : | Este estudo teve como objetivo avaliar possíveis desvios na planura e simetria do feixe de fótons de um acelerador linear Precise, da empresa Elekta, instalado na Fundação de Beneficência Hospital de Cirurgia (FBHC). Devido à indisponibilidade do sistema de varredura automática, os testes foram realizados manualmente, por meio de medições ponto a ponto com uma câmara de ionização imersa em um objeto simulador preenchido com água. Foram analisados campos de 10 × 10 cm² e 30 × 30 cm², com energias de 6 MeV e 10 MeV, seguindo as diretrizes do manual do fabricante (Elekta) e as recomendações da Agência Internacional de Energia Atômica (IAEA). Os resultados foram comparados com os dados simulados pelo sistema de planejamento de tratamento MONACO V6.3.3, com o intuito de verificar a confiabilidade dos parâmetros clínicos. As medições permitiram a construção dos perfis de dose, por meio dos quais foi possível avaliar a uniformidade do feixe. Os resultados demonstraram que não há desvios significativos de planura e simetria, estando todos os valores dentro dos limites de tolerância recomendados, tanto pelos testes baseados no manual da Elekta quanto pelas orientações da IAEA. Adicionalmente, os testes de penetração indicaram que o feixe mantém adequadas condições de profundidade, essenciais para garantir a eficácia terapêutica. No entanto, foram observadas discrepâncias entre os resultados medidos e os gerados pelo sistema de planejamento, sugerindo a necessidade de aprimoramentos técnicos, como o uso de imagens tomográficas com cortes mais finos, para maior precisão nos cálculos de dose. Apesar disso, tais variações não comprometeram a conformidade do perfil do feixe, que permaneceu dentro dos padrões aceitáveis. Conclui-se, portanto, que a distribuição de dose entregue ao volumealvo está de acordo com o planejado, assegurando a qualidade e a segurança do tratamento radioterápico. |
Abstract: | This study aimed to evaluate potential deviations in the flatness and symmetry of the photon beam produced by a Precise linear accelerator manufactured by Elekta, installed at the Fundação de Beneficência Hospital de Cirurgia (FBHC). Due to the unavailability of the automatic scanning system, tests were performed manually by measuring dose point by point using an ionization chamber immersed in a water-filled phantom. Fields of 10 × 10 cm² and 30 × 30 cm² were analyzed, with beam energies of 6 MeV and 10 MeV, following the guidelines of the manufacturer’s manual (Elekta) and the recommendations of the International Atomic Energy Agency (IAEA). The measured results were compared with those generated by the MONACO V6.3.3 treatment planning system, in order to ensure the reliability of the clinical parameters. The measurements allowed for the construction of dose profiles, through which beam uniformity was evaluated. The results showed no significant deviations in flatness and symmetry, with all values remaining within the recommended tolerance limits, according to both the Elekta manual and IAEA guidelines. Additionally, beam penetration quality tests confirmed that the beam maintains adequate depth-dose characteristics, which are essential for therapeutic effectiveness. However, discrepancies were observed between the measured data and the treatment planning system results, indicating a need for technical improvements, such as the use of CT images with thinner slices, to increase dose calculation accuracy. Nevertheless, these differences did not compromise the beam profile, which remained within acceptable limits. It is therefore concluded that the dose distribution delivered to the target volume is consistent with the planned values, ensuring the quality and safety of radiotherapy treatments. |
Keywords: | Física Física médica Ensino superior (UFS) Radiologia Fotobiologia Acelerador linear (LINAC) Garantia da qualidade dosimétrica Radioterapia Beam flatness Beam symmetry Linear accelerator (LINAC) Dosimetric quality assurance Radiotherapy Fundação de Beneficência Hospital de Cirurgia (FBHC) |
Subject CNPQ: | CIENCIAS BIOLOGICAS::BIOFISICA::RADIOLOGIA E FOTOBIOLOGIA |
Language: | por |
Institution: | Universidade Federal de Sergipe (UFS) |
Department: | DFI - Departamento de Física – Física Médica - São Cristóvão - Presencial |
Citation: | Silva, Rayanna Vieira da. Estudo do perfil de feixe de fótons em LINAC. São Cristóvão, 2025. Monografia (graduação em Física Médica) – Departamento de Física, Centro de Ciências Exatas e Tecnologia, Universidade Federal de Sergipe, São Cristóvão, SE, 2025 |
URI: | https://ri.ufs.br/jspui/handle/riufs/22482 |
Appears in Collections: | Física |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Rayanna_Vieira_Silva.pdf | 1,89 MB | Adobe PDF | ![]() View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.