Please use this identifier to cite or link to this item: https://ri.ufs.br/jspui/handle/riufs/24075
Document Type: Artigo
Title: Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor
Authors: Pupim, Denise
Iwaki Filho, Liogi
Takeshita, Wilton Mitsunari
Iwaki, Lilian Cristina Vessoni
Issue Date: 2013
Resumo : Context: The pulse oximeter is a device that noninvasively provides continuous information about the peripheral oxygen saturation (SpO2) rate. This device is utilized in the detection of hypoxemia, due to its able to sense changes in hemoglobin oxygen saturation. Aims: The objective of this study was to verify the accuracy of the Choice® Medical MD300C3 Fingertip Pulse Oximeter, as compared to that of a hospital oximeter coupled with a Dräger® Infinity Delta monitor, with the purpose of using this first methodology in dental procedures to monitor the peripheral oxygen saturation (SpO2) of patients submitted to dental treatments. Materials and Methods: Fifty‑five adult patients, both genders, were selected in the Santa Casa Hospital of Maringa, Brazil. The volunteers did not present cardiac problems, prosthetic cardiac valves, pacemakers, or pulmonary diseases, and were not pregnant or children. Each patient received a portable fingertip pulse oximeter (PPO) on the middle finger of the left hand and the hospital oximeter (control device) on the forefinger of the same hand. A total of six measurements were developed. The Pearson correlation coefficient and the Bland and Altman method was used to calculate the statistical analysis. Results: No statistically significant difference was found between the measurements taken by the utilized devices. The average of comparative analysis presented by the devices was 0.2337 ± 0.4355 (mean ± SD), suggesting a strong correlation between the obtained results. Conclusion: According to the methodology of the research, the PPO has similar accuracy to the conventional hospital oximeter with digital sensor. The PPO can be used in dental treatments.
Keywords: Dentistry
Oral medicine
Oximetry
Pulse oximetry
Surgery
Vital signs
ISSN: 1998-3603
Is part of: Indian Journal of Dental Research
Language: eng
Publisher / Institution : Medknow Publications
Citation: PUPIM, D. et al. Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor. Indian Journal of Dental Research, Mumbai, v. 24, n. 5, p. 542-546, 2013. Disponível em: https://journals.lww.com/ijdr/fulltext/2013/24050/evaluation_of_accuracy_of_portable_fingertip_pulse.6.aspx. Acesso em: 15 dez. 2025.
License: Creative Commons Atribuição-NãoComercial-CompartilhaIgual 4.0 Internacional (CC BY-NC-SA 4.0)
URI: https://ri.ufs.br/jspui/handle/riufs/24075
Appears in Collections:DOD - Artigos de periódicos

Files in This Item:
File Description SizeFormat 
EvaluationAccuracyPortableFingertipPulseOximeter.pdf548,63 kBAdobe PDFThumbnail