Use este identificador para citar ou linkar para este item: https://ri.ufs.br/jspui/handle/riufs/24075
Registro completo de metadados
Campo DCValorIdioma
dc.contributor.authorPupim, Denise-
dc.contributor.authorIwaki Filho, Liogi-
dc.contributor.authorTakeshita, Wilton Mitsunari-
dc.contributor.authorIwaki, Lilian Cristina Vessoni-
dc.date.accessioned2025-12-15T18:16:05Z-
dc.date.available2025-12-15T18:16:05Z-
dc.date.issued2013-
dc.identifier.citationPUPIM, D. et al. Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor. Indian Journal of Dental Research, Mumbai, v. 24, n. 5, p. 542-546, 2013. Disponível em: https://journals.lww.com/ijdr/fulltext/2013/24050/evaluation_of_accuracy_of_portable_fingertip_pulse.6.aspx. Acesso em: 15 dez. 2025.pt_BR
dc.identifier.issn1998-3603-
dc.identifier.urihttps://ri.ufs.br/jspui/handle/riufs/24075-
dc.languageengpt_BR
dc.publisherMedknow Publicationspt_BR
dc.relation.ispartofIndian Journal of Dental Researchpt_BR
dc.subjectDentistryeng
dc.subjectOral medicineeng
dc.subjectOximetryeng
dc.subjectPulse oximetryeng
dc.subjectSurgeryeng
dc.subjectVital signseng
dc.titleEvaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensorpt_BR
dc.typeArtigopt_BR
dc.identifier.licenseCreative Commons Atribuição-NãoComercial-CompartilhaIgual 4.0 Internacional (CC BY-NC-SA 4.0)pt_BR
dc.description.resumoContext: The pulse oximeter is a device that noninvasively provides continuous information about the peripheral oxygen saturation (SpO2) rate. This device is utilized in the detection of hypoxemia, due to its able to sense changes in hemoglobin oxygen saturation. Aims: The objective of this study was to verify the accuracy of the Choice® Medical MD300C3 Fingertip Pulse Oximeter, as compared to that of a hospital oximeter coupled with a Dräger® Infinity Delta monitor, with the purpose of using this first methodology in dental procedures to monitor the peripheral oxygen saturation (SpO2) of patients submitted to dental treatments. Materials and Methods: Fifty‑five adult patients, both genders, were selected in the Santa Casa Hospital of Maringa, Brazil. The volunteers did not present cardiac problems, prosthetic cardiac valves, pacemakers, or pulmonary diseases, and were not pregnant or children. Each patient received a portable fingertip pulse oximeter (PPO) on the middle finger of the left hand and the hospital oximeter (control device) on the forefinger of the same hand. A total of six measurements were developed. The Pearson correlation coefficient and the Bland and Altman method was used to calculate the statistical analysis. Results: No statistically significant difference was found between the measurements taken by the utilized devices. The average of comparative analysis presented by the devices was 0.2337 ± 0.4355 (mean ± SD), suggesting a strong correlation between the obtained results. Conclusion: According to the methodology of the research, the PPO has similar accuracy to the conventional hospital oximeter with digital sensor. The PPO can be used in dental treatments.pt_BR
dc.description.localMumbaipt_BR
Aparece nas coleções:DOD - Artigos de periódicos

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
EvaluationAccuracyPortableFingertipPulseOximeter.pdf548,63 kBAdobe PDFThumbnail
Visualizar/Abrir


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.