Data do documento: 2016
Resumo: In this paper, various simplifications of the HDR source Varian VariSource Classic model, in which 192Ir as a radionuclide is used, were compared. These simplifications were carried out by Monte Carlo simulations, using the MCNPX 2.7.0 code. The different sources were compared through a distribution of energy deposition in a water phantom. Our results indicated that small simplifications will present no influence on the source response, and the removal of the entire capsule surrounding the radionuclide will present a difference of just 0.53% in the final response.
Palavras-chave: HDR source
VariSource classic model
Monte Carlo simulations
192Ir sources
ISSN: 1742-6596
Parte de : Journal of Physics: Conference Series
Idioma: eng
Instituição/Editora: IOP Publishing
Citação: SANTOS, W. S. et al. The influence of different 192Ir sources geometries to the energy deposition. Journal of Physics: Conference Series, Bento Gonçalves, v. 733, nov./dez. 2016. Disponível em: https://iopscience.iop.org/article/10.1088/1742-6596/733/1/012092. Acesso em: 21 jul. 2026.
Licença: Creative Commons Atribuição 3.0 Não Adaptada (CC BY 3.0)
Identificador: https://doi.org/10.1088/1742-6596/733/1/012092
URI: https://ri.ufs.br/jspui/handle/riufs/25622
Aparece nas coleções:DFI - Artigos de periódicos

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
InfluenceDifferent192IrSource.pdf729,82 kBAdobe PDFThumbnail
Visualizar/Abrir


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.